Abstract:

A 69-year-old woman with sick sinus syndrome underwent permanent pacemaker implantation in our hospital. A mild and persistent chest pain occurred 2 days after device implantation, and myocardial perforation without pericardial effusion was found at the tip of the right ventricular lead by two-dimensional transthoracic echocardiography. The diagnosis was then confirmed by fluoroscopy, and the lead was repositioned under fluoroscopic guidance.

Key words: cardiac pacing; myocardial perforation; echocardiography